Some electrodes are showing as yellow and are blocked at 5%.

Some electrodes are showing as yellow and are blocked at 5%.

This means that the system has identified an open circuit and, for your safety, is blocking stimulation.

Take the following measures to try and resolve the situation:;
1. Ensure that the User is not wearing anything under the Neuro20 Smart Suit which could obstruct proper connection of the electrodes with the skin ... i.e. undergarments or bra.;
2. Ensure Users are in the proper size suit (ie. not too bulky suit).
3. Ensure that there is moisture on the electrode – either from sweat or use a spray bottle to wet the electrodes.
4. If the indication persists, then remove and re-clip the Neuro20 Control Box.
5. If you have a different Neuro20 Control Box available to you, then try using the other Control Box.
6. If the indication persists, then close and restart the Software.
7. If the indication persists, then change the Neuro20 Smart Suit.